Error messages for all models
|
|
MEANING
|
POSSIBLE SOLUTIONS
|
ERR01 |
Weight not stable after operation of tare |
Protect the balance from air flows or from vibrations of the working table
|
ERR02 |
Impossible to start the calibration due to instability of the balance |
Protect the balance from air flows or from vibrations of the working table
|
ERR03 |
Calibration weight not correct or balance unstable |
Calibrate with correct weight or protect the balance from environment disturbs
|
ERR04 |
Weight of samples for the piececounting function not adequate or unstable. |
Select a bigger number of samples or protect the balance from vibrations.
|
ERR05 |
Impossible to print / weight unstable. |
Protect the balance from environment disturbs.
|
ERR06 |
Weight cannot get stable in density mode |
Protect the balance from environment disturbs.
|
ERR07 |
Weight cannot get stable in percentual weighing mode |
Protect the balance from environment disturbs.
|
ERR08 |
Anomaly on autocalibration motor. |
|
ERR09 |
Weight cannot get stable in formulation mode |
Protect the balance from environment disturbs
|
ERR10 |
Weight of component out of tolerance in formulation mode |
|
ERR F |
Weight of component out of tolerance in formulation mode |
Ask assistance to authorized service staff.
|
CAL BUT |
The balance requires to be recalibrated. |
Unload weights, if any, on the pan, and press the CAL button.
|
Error messages for i-Thermo moisture analyzer
|
|
MEANING
|
POSSIBLE SOLUTIONS |
ERR07 |
Error in inserting the data
|
|
ERR 10 |
The weight does not reach stability before start of dry
|
Protect the balance from environment disturbs |
ERR 11 |
Weight of substance not enough
|
Increase quantity of substance |
Error Message on Displays
|
Over-range condition
|
Unload the weights loaded on the pan |
|
Under-range condition |
Place properly pan and underpan
|